Cieslak, Michal and Akurugoda Gamage, Kelum and Glover, Robert (2017) Investigation into a suitable scintillator and coded-aperture material for a mixed-field radiation imaging system. Journal of Instrumentation, 12 (12): P12007. ISSN 1748-0221
AAManuscript.pdf - Accepted Version
Available under License Creative Commons Attribution-NonCommercial.
Download (9MB)
Abstract
Monte-Carlo modelling (MCNPX) methods have been employed to conduct an investigation into a suitable scintillator and coded-aperture material for a scintillator based mixed-field radiation imaging system. Single stilbene crystal, pure and 6Li-loaded plastic scintillators were simulated and their neutron/gamma detection performance compared when exposed to the spontaneous fission spectrum produced by 252Cf. The most suitable candidate was then incorporated into a scintillator based mixed-field coded-aperture imaging system. Coded-aperture models made of three W and 113Cd compositions were tested in different neutron/gamma environments with a square W collimator modelled around the aperture. Each simulation involved recording the interactions of neutron events in organic solid scintillator, whose neutron/gamma detection performance was assessed prior to the coded-aperture material investigation. Three coded-aperture material compositions have been tested with the simulated 252Cf spontaneous fission as well as 241AmBe neutron sources. Results generally claim very good detection sensitivity and spatial resolution for the radioactive sources located in the centre of the aperture.