Dual-mode terahertz time-domain spectroscopy system

Ung, B. S. Y. and Li, Jining and Lin, Hungyen and Fischer, B. M. and Withayachumnankul, W. and Abbott, D. (2013) Dual-mode terahertz time-domain spectroscopy system. IEEE Transactions on Terahertz Science and Technology, 3 (2). pp. 216-220. ISSN 2156-342X

Full text not available from this repository.

Abstract

Terahertz time-domain spectroscopy (THz-TDS) systems traditionally operate in a single mode, either in reflection or transmission. In cases where the sample has nonunity permeability, measurements in both reflection and transmission geometries are required. The process of shifting and swapping the samples during an experiment increases the measurement uncertainty. This paper therefore presents a system where both reflection and transmission measurements can be performed simultaneously to reduce both experimental error and acquisition time. The measurement results are validated against findings in literature.

Item Type:
Journal Article
Journal or Publication Title:
IEEE Transactions on Terahertz Science and Technology
Uncontrolled Keywords:
/dk/atira/pure/subjectarea/asjc/3100/3108
Subjects:
?? radiationelectrical and electronic engineering ??
ID Code:
76816
Deposited By:
Deposited On:
23 Nov 2015 15:26
Refereed?:
Yes
Published?:
Published
Last Modified: