The effect of aluminium coating on elemental signals in X-ray microanalysis.

Reid, A. P. and Oates, K. and Potts, W. T. W. (1993) The effect of aluminium coating on elemental signals in X-ray microanalysis. Microscopy Research and Technique, 24 (2). pp. 168-172. ISSN 1059-910X

Full text not available from this repository.

Abstract

It has been determined that, in the normal range of aluminium coating thicknesses used to remove charge from non-conducting specimens in the electron microscope, no detectable influence on the elemental signals obtained in X-ray microanalysis is observed. This is in contrast to a previous report (Hopkins et al., J. Electron Microsc. Tech., 18:176-182, 1991) of a reduction in elemental signal with increasing aluminium coating thickness. An explanation of errors in the previous interpretation is provided.

Item Type:
Journal Article
Journal or Publication Title:
Microscopy Research and Technique
Uncontrolled Keywords:
/dk/atira/pure/researchoutput/libraryofcongress/ge
Subjects:
?? CHARGE DISSIPATION • ABSORPTION • ATTENUATION • ELECTRON BEAMANATOMYINSTRUMENTATIONMEDICAL LABORATORY TECHNOLOGYHISTOLOGYGE ENVIRONMENTAL SCIENCES ??
ID Code:
22679
Deposited By:
Deposited On:
20 Jan 2009 09:16
Refereed?:
No
Published?:
Published
Last Modified: