Polymeric biomaterial wear test rig validated to ASTM F732-00 and against clinical data

Joyce, T. J. and Monk, D. and Thompson, P. and Chiu, P. and Unsworth, A. and Green, S. M. (2004) Polymeric biomaterial wear test rig validated to ASTM F732-00 and against clinical data. In: Transactions - 7th World Biomaterials Congress. Transactions - 7th World Biomaterials Congress . UNSPECIFIED, AUS. ISBN 1877040193

Full text not available from this repository.

Abstract

The test of three biopolymers were conducted using a modified pin-on-plate rig to evaluate whether the machine could be validated to the ASTM standard. The modification entailed the addition of rotational motion of the test pins, to give multi-directional motion. Pins were machined to form flat ended cylinders and articulated against stainless steel plates. The results show that the biopolymers subjected to multi-directional motion exhibited relevant values of wear and ASTM recommended targets were achieved.