Automated Response Matching for Organic Scintillation Detector Arrays

Aspinall, Michael and Joyce, M.J. and Cave, F.D. and Plenteda, R. and Tomanin, A. (2017) Automated Response Matching for Organic Scintillation Detector Arrays. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 861. pp. 46-54. ISSN 0168-9002

[thumbnail of 1-s2.0-S0168900217305338-main]
Preview
PDF (1-s2.0-S0168900217305338-main)
1_s2.0_S0168900217305338_main.pdf - Accepted Version
Available under License Creative Commons Attribution-NonCommercial-NoDerivs.

Download (707kB)

Abstract

This paper identifies a digitizer technology with unique features that facilitates feedback control for the realization of a software-based technique for automatically calibrating detector responses. Three such auto-calibration techniques have been developed and are described along with an explanation of the main configuration settings and potential pitfalls. Automating this process increases repeatability, simplifies user operation, enables remote and periodic system calibration where consistency across detectors’ responses are critical.

Item Type:
Journal Article
Journal or Publication Title:
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Additional Information:
This is the author’s version of a work that was accepted for publication in Physics Reports. Changes resulting from the publishing process, such as peer review, editing, corrections, structural formatting, and other quality control mechanisms may not be reflected in this document. Changes may have been made to this work since it was submitted for publication. A definitive version was subsequently published in Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 861, 2017 DOI: 10.1016/j.nima.2017.05.004
Uncontrolled Keywords:
/dk/atira/pure/subjectarea/asjc/3100/3105
Subjects:
?? detector response matchingreal time systemsfast neutron detectionradiation monitoringspectroscopydigital signal processinginstrumentationnuclear and high energy physics ??
ID Code:
86261
Deposited By:
Deposited On:
15 May 2017 08:18
Refereed?:
Yes
Published?:
Published
Last Modified:
31 Dec 2023 00:48